Handyscope HS3 Hardware specifications
Acquisition system
Number of input channels 2 analog, BNC
    Resolution 8 bit @ 100 MHz
12 bit @ 50 MHz
14 bit @ 3.125 MHz
16 bit @ 195 kHz
    Accuracy 0.2% ± 1 LSB
    Sensitivity 200 mVolt .. 80 Volt full scale
    Coupling AC/DC
    Impedance 1 MOhm / 30 pF
    Protection 200 Volt (DC + AC peak < 10 kHz)
    Bandwidth (-3dB) DC to 50 MHz maximum
    AC coupling cut off frequency (-3dB) 1.5 Hz
Maximum sampling rate 100 MHz,   10 nsec (model HS3-100)
50 MHz,   20 nsec (model HS3-50  )
25 MHz,   40 nsec (model HS3-25  )
10 MHz, 100 nsec (model HS3-10  )
5 MHz, 200 nsec (model HS3-5    )
Sampling source internal ±0.01%, external
Memory 131072 samples each channel
 
Arbitrary Waveform Generator (independent from acquisition system)
Number of output channels 1 analog, BNC
    Resolution 14 bit @ 50 MHz
    Amplitude -12 Volt .. 12 Volt
    Amplitude step 0 - ±0.1 V range, 8192 steps
±0.1 - ±0.9 V range, 8192 steps
±0.9 - ±12 V range, 8192 steps
    Coupling DC
    Impedance 50 Ohm
    Bandwidth DC to 2MHz
    DC level 0 - ±12 V in 8192 steps
Sampling rate 50 MHz
Sampling source internal, 0.01%
Memory 1024 points DDS mode
128K points linear mode
Waveforms sine, triangle, square, DC, noise and user defined
Symmetry 1 - 99%, 1% steps
 
Trigger system
System digital, 2 levels
Source CH1, CH2, AND, OR, digital external
Trigger modes rising slope, falling slope, inside window, outside window, peak
Level adjustment 0 - 100% of full scale
Hysteresis adjustment 0 - 100% of full scale
Resolution 0.025% (12 bits)
Digital trigger input level 0 - 3.3 volt (5 volt max)
Pre trigger 0 - 131071 samples (0 - 100%, one sample resolution)
Post trigger 0 - 131071 samples (0 - 100%, one sample resolution)
 
Interface
Interface USB 2.0 (USB 1 also supported)
 
Power Requirements
Power from USB port 500 mA max (2.5 Watt max)
Power via extension connector 1500 mA max (7.5 Watt max)
 
Physical
Instrument height 25 mm (1 inch)
Instrument length 170 mm (6.7 inch)
Instrument width 140 mm (5.2 inch)
Cord length 1.8 meter (70 inch)
Weight 480 gram (17 ounce)
Accessories 2 oscilloscope probes 1:1 - 1:10 switchable
external power cable for PS/2 mouse or keyboard connector
 
Options
Optionally available 10 MHz generator output
Optionally available Single channel Handyscope HS3 without generator, but with 512K samples memory